9 Obour buildings - Salah Salem
01011929006 egymst@egymst.com
Close
9 Obour buildings - Salah Salem 01011929006
+20 24031849 egymst@egymst.com
Hours: Sun - Thu 9:00 AM - 5:00 PM

Cygnus Mini ROV Mountable

Cygnus Instruments Ltd

The Cygnus Mini ROV Mountable thickness gauge has been specially designed to be mounted onto small observation ROVs with or without a manipulator facility. Built in conjunction with VideoRay, this subsea thickness gauge will allow the operator to carry out an ultrasonic thickness measurement in areas that have previously been unfeasible using larger ROVs or divers. And no need to remove coatings! Ideal for use in UWILD or IWS class surveys, subsea inspection, repair and maintenance (IRM) of infrastructure and underwater equipment.

Contact Us

Description

Key Features:

  • Multiple-Echo mode for accurate, through-coat measurements as specified by Classification Societies

  • Supplied with CygLink software to display and process measurements on a computer at the surface

  • Easy calibration at the surface via CygLink software

  • Optional purpose-built ROV probe handlers for optimal alignment.

Materials

Velocities between 1,000 and 9,995 m/s (0.0400 and 0.3998 in/microsec)

Measurement Range in Steel

1 – 250 mm (0.040 – 10 inch) depending on selected probe and configuration, material and temperature

Accuracy

±0.1 mm (±0.004 inch) or 0.1% of thickness measurement, whichever is greatest, when calibrated in accordance with Cygnus Instruments calibration procedure

Resolution

0.1 or 0.05 mm (selectable) (0.005 or 0.002 inch)

Probe Options

Single crystal probes

Power

7.5 – 30 V DC @ 150 mA (max)

Display

PC or laptop with VGA (not included)

Size

160 x 62 mm (6.3 x 2.4 inch)

Weight in Air

550 g (19.4 oz)

Operating Temp.

-10°C to +50°C (14°F to 122°F)

Approvals

RINA Type approved

Testing

Tested to 500 m (1,640 ft) depth

Communication

RS-422, Simplex Single Pair or 2400 Baud (RS-232 9600 Baud output available as special order)

Standards

Designed for BS EN ISO 16831

Compliance

CE, UKCA, RoHS

Warranty

3 years on gauge and 6 months on probe